Shanghai Haozhuang has a well-recognized team of experts in the industry and first-class scientific and technological talents. Shanghai Haozhuang Instrument Co., Ltd. has developed a small spray dryer of LNB brand, which is energy-saving and environmentally friendly, leading the country.

Installation instructions: 1. Installation of the drying chamber Hold the drying chamber with both hands and insert it into the fixing clamp of the drying chamber obliquely, and lock the lock nut of the drying chamber (subject to the rotation of the drying chamber in the clamp).
2. Installation of the cyclone separator Put the lock nut, seal ring and stainless steel gasket of the cyclone separator on the air outlet pipe of the cyclone separator, and then insert it into the air outlet pipe of the equipment together to adjust the air outlet of the drying chamber and the cyclone separator Position the air inlet so that the two ports are aligned straight, connect the two ports with a clamp, and finally lock the lock nut of the cyclone separator.
3. Connect the collection bottle and cyclone separator with a clamp 4. Connect the collection tube to the drying chamber with a clamp 5. Install the spray chamber to the equipment, connect the 4mm gas tube (for needle) and 6mm gas tube (For spray)
6. Install the food-grade silicone tube on the peristaltic pump and insert it into the feed port of the spray chamber. Instructions for peristaltic pump click "open", peristaltic pump "start", and then click "stop" to close. Automatically control the operation of the peristaltic pump, click The blue area of ​​the automatic button of the peristaltic pump (below) (the button changes to "off"). The peristaltic pump starts automatically. The speed of the peristaltic pump can be automatically adjusted according to the set value of the outlet temperature, so that the feed rate of the peristaltic pump is clicked. For manual operation, "Automatic" is displayed. [1] The fan controls the start and stop of the fan, click the fan button area (upper), the fan starts (the button moves to "start"), click the lower, the fan stops (the button moves to "stop" ") .. The needle is used to control the start and stop of the needle, click on the black area (above) of the pass button, start (the button moves to" start "), click below, stop (the button moves to" stop "), pass the needle The operating speed is changed in the interface (3) by changing the setting value of the needle setting. The air compressor controls the start and stop of the air compressor, click the air compressor button black area (above), the air compressor start (The button moves to "Start"), click below, the air compressor stops (the button moves to "Stop"). The heater controls the start and stop of the electric heater, click the heater button black area (above), the heater starts ( The button moves to "Start"), click below, the heater stops (the button moves to "Stop"). Note: The heater will not start before the fan is not started, turn off the fan heater and automatically disconnect. The fan settings are set Set the rotation frequency of the extension, press the value box to pop up the numeric keyboard, press the CLR key to clear the number, and then enter the desired value, press the ENTER key to complete the modification. The peristaltic pump manually controls the operation of the peristaltic pump, click the peristaltic pump automatically Button black area (upper) (display ON), the peristaltic pump starts automatically, the lower display, the peristaltic pump stops (display OFF) through the needle to set the operating frequency through the needle, the value represents a few seconds to start, press the value box, The numeric keyboard pops up, press the CLR key to clear the number, and then enter the desired value, press the ENTER key to modify.
Peristaltic pump setting Set the feed rate of the peristaltic pump (r / m), press the value box to pop up the numeric keyboard, press the CLR key to clear the number, then enter the desired value, press the ENTER key to modify. Description: Click the control panel to enter the interface (2), click the main menu to enter the interface (1) The inlet air temperature is automatically adjusted to adjust the stability of the inlet air temperature, if and only when it is dry, the inlet air temperature display value and inlet air temperature setting are found When there is a large fluctuation between the values, the inlet air temperature self-adjustment is required to ensure that the inlet air temperature setting value is consistent with the inlet air temperature display value, and the inlet air temperature setting value and inlet air must be The temperature difference between the temperature display values ​​is more than 50 degrees. For example, if the inlet air temperature is set at 120 degrees, the inlet air temperature is displayed at 70 degrees, that is, the set value> the displayed value is more than 50 degrees. Click the black zone (upper) (display ON) of the inlet air temperature self-adjustment button, and the inlet air temperature self-adjustment starts. After the self-discussion is completed, it will automatically end. If the self-adjustment is completed before the self-adjustment is completed, you need to click the end button of the inlet air temperature self-adjustment. Note: After the fan and heater are started, the inlet air temperature self-adjusting function is set. The inlet air temperature setting value sets the inlet air temperature. Press the value box to pop up the numeric keyboard, press the CLR key to clear the number, and then enter the desired , Press ENTER to modify. The inlet air temperature display value shows the actual value of the inlet air temperature. The end of the inlet air temperature self-adjustment is blocked. During the process of the inlet air temperature self-adjustment, if you want to interrupt the self-adjustment halfway, press this button. When the room finds that there is a relatively large fluctuation between the displayed value of the outlet temperature and the set value of the outlet temperature, the outlet temperature needs to be adjusted automatically to ensure that the setting value of the outlet temperature is consistent with the displayed value of the outlet temperature, open the note money It must be ensured that there is a temperature difference of more than 50 degrees between the set value of the outlet temperature and the displayed value of the outlet temperature. For example, if the outlet temperature is set at 70 degrees, the displayed value of the outlet temperature is above 120 degrees, that is, the display value> the set value Above 50 degrees. Set the outlet temperature setting value, press the value box to pop up the numeric keyboard, press the CLR key to clear the number, then enter the desired value, press the ENTER key to modify. The display value of the outlet temperature shows the actual value of the outlet temperature. When the outlet temperature self-adjustment is finished, if you want to interrupt the self-adjustment in the middle of the outlet temperature self-adjustment, press this button to eliminate common problems. 1. The fan does not work. Damaged, the fan is damaged, contact our company 2. The electric heater does not work, the fan does not start, the solid state relay is damaged, the electric heater is damaged, start the fan, replace the solid relay, contact my company, 3. The air compressor does not work, the intermediate relay is damaged, the air compressor does not start If the air compressor is damaged, replace the intermediate relay to start the air compressor. 4. The device has no power. The external socket is unreliable. Air switch 1 is in the closed position. The intermediate relay is damaged. Check whether the external power supply has power. Turn on the air switch 1. Replace the intermediate relay. The start button on the panel is damaged, the start button is replaced, the 24V switching power supply is damaged, the touch screen is damaged, replace the switching power supply, please contact our company. 6. The PLC does not work. The air switch 2 is in the closed position.
7. Inlet air temperature is not displayed, PT-100 temperature probe is loosely connected, PT-100 is damaged, PT temperature module is damaged, please contact our company 8. Outlet air temperature is not displayed, PT-100 temperature probe is loosely connected, PT-100 is damaged, PT temperature The module is damaged. Contact our company. 9. The inlet air temperature cannot reach the set value. The fan air volume is too large. Modify the fan parameters. 10. The temperature of the drip feed air at the bottom of the drying chamber is too low. The atomizing air pressure is too low. The compressed air leakage air feed volume Increase the inlet air temperature too much. After opening the device, adjust the pressure to 2-3 Bar (φ6 white air pipe). Check whether there is air leakage at each connection. Modify the peristaltic pump parameters. 11. The needle is not working. The air valve is not open. The pressure is too small. If the solenoid valve is too large, the valve will be damaged. Open the valve (φ4 white trachea). Increase the pressure of the pressure reducing valve. Modify the needle parameters. Replace the solenoid valve. Seek credibility, so choose because of quality, we promise to be a product that customers can rest assured!